Calibration of a Six-Axis Serial Robotic Arm Mounted to a Moveable Base

Michael Allman, David R. Novotny, Joshua A. Gordon

Title graphic for CMSJ-13-2-01: Calibration of a Six-Axis Serial Robotic Arm Mounted to a Moveable Base

Abstract

In this work, we present the calibration of a serial robotic positioner mounted to a rail system. This robot is one of two positioners that make up the Large Antenna Positioning System (LAPS) in the Communications Technology Laboratory (CTL) at the National Institute of Standards and Technology (NIST). The out-of-box positioning accuracy is approximately 1 mm. Using a custom Python package developed at NIST for automation, analysis, and control, we calibrated the robot’s kinematic model, improving the positioning accuracy to approximately 0.2 mm over a total working volume of approximately 10 m3. We discuss effects detrimental to accuracy, such as mechanical backlash (hysteresis) and spatial drift, and methods to mitigate them. INTRODUCTION CTL developed LAPS to support over-the-air (OTA) testing needs for modern communications systems as well as perform general antenna characterization in the commonly used bands between 0.5 GHz and 30 GHz. The performance of any antenna test facility depends on the ability to accurately position source and probe antennas over a scanning volume.