Abstract
In this work, we present the calibration of a serial robotic positioner mounted to a rail system. This robot is one of two positioners that make up the Large Antenna Positioning System (LAPS) in the Communications Technology Laboratory (CTL) at the National Institute of Standards and Technology (NIST). The out-of-box positioning accuracy is approximately 1 mm. Using a custom Python packaged developed at NIST for automation, analysis, and control, we calibrated the robot’s kinematic model, improving the positioning accuracy to approximately 0.2 mm over a total working volume of approximately 10 m3. We discuss effects detrimental to accuracy such as mechanical backlash (hysteresis) and spatial drift and methods to mitigate them.
Also published as CMSJ-13-2-01, The Journal of the CMSC, Volume 13, Number 2 (Autumn 2018).
Cite this paper
Michael Allman, David Novotny, Josh Gordon. National Institute of Standards and Technology (NIST), Communications Technology Laboratory, Boulder, CO. Presented at the Coordinate Metrology Systems Conference 2018, Reno, Nevada, July 23-27, 2018. Coordinate Metrology Society, reference CMS-2018-020.